Spatial sensitivity and penetration depth of three cerebral oxygenation monitors
Sonny Gunadi1, Terence S Leung1, Clare E Elwell1
1Department of Medical Physics and Biomedical Engineering, University College London, Gower Street, London, WC1E 6BT, UK.
Abstract:
The spatial sensitivities of NIRO-100, ISS Oximeter and TRS-20 cerebral oxygenation monitors are mapped using the local perturbation method to inform on their penetration depths and susceptibilities to superficial contaminations. The results show that TRS-20 has the deepest mean penetration depth and is less sensitive than the other monitors to a localized absorption change in the superficial layer. However, an integration time of more than five seconds is required by the TRS-20 to achieve an acceptable level of signal-to-noise ratio, which is the poorest amongst the monitors. With the exception of NIRO-100 continuous wave method, the monitors are not significantly responsive to layer-wide absorption change that occurs in the superficial layer.