Related Experiment Video
Updated: Apr 20, 2026

Measurement of X-ray Beam Coherence along Multiple Directions Using 2-D Checkerboard Phase Grating
Published on: October 11, 2016
Simultaneous thickness and group index measurement with a single arm low-coherence interferometer
Abstract:
We present a single arm low-coherence interferometer to directly measure the physical thickness and group refractive index of optically transparent samples having flat and parallel surfaces. The optical arrangement, resembling a common-path interferometer, is more compact and stable than the usual dual-arm low-coherence interferometer. It has been used to measure samples of Herasil 102 fused silica, Schott B270 Superwhite crown glass and borosilicate cover glass. The results obtained indicate uncertainties in the third decimal place for index values and thicknesses accurate to within 2 μm.
More Related Videos
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Interference and Diffraction
Confocal Fluorescence Microscopy