Comparison of total haemoglobin mass measured with the optimized carbon monoxide rebreathing method across different
Abstract:
A new Radiometer™ hemoximeter (ABL-80) has recently become available to measure carboxyhaemoglobin concentration for the optimized CO-rebreathing method (oCOR-method). Within the English Institute of Sport (EIS), hemoximeters are used in three different laboratories; therefore, precision and agreement of total haemoglobin mass (tHbmass) determination across sites is essential, and comparison to the previous model OSM-3 is desirable. Six male and one female (age 30 ± 6 years, body mass 78.1 ± 10.6 kg) undertook the oCOR-method. Venous blood (~2 ml) was sampled immediately before and at 7 min during the oCOR-method; with seven replicates from each time point simultaneously analysed on five different Radiometer™ hemoximeters [OSM-3(1), OSM-3(2), ABL-80(1), ABL-80(2) and ABL-80(3)]. There were no differences (p > 0.05) between Δ%HbCO or mean tHbmass analysed with five different hemoximeters (OSM-3(1): 886 ± 167 g; OSM-3(2): 896 ± 160 g: ABL-80(1): 904 ± 157 g; ABL-80(2): 906 ± 163 g: ABL-80(3): 906 ± 162 g). However, the Bland-Altman plot revealed that there was closer agreement between ABL-80 machines for tHbmass than for the OSM-3. The variance (i.e. % error) across replicate samples decreased as the number of samples increased, with the error derived from the 'worse-case' scenario (single samples) being 1.2 to 1.6 fold greater in the OSM-3 than the ABL-80. Although there were no differences in the average tHbmass measured with five different hemoximeters, the new ABL-80 were in better agreement with each other compared to the old OSM-3. Previously, five replicates were required to achieve a low error using the OSM-3; however, three replicates are sufficient with the ABL-80 model to produce an error of ≤ 1% in tHbmass.
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Transport in the Blood
Hemoglobin
When all four heme groups are bound to oxygen, the resulting molecule is called oxyhemoglobin. As a result, arterial blood...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...