Related Experiment Video
Updated: Apr 20, 2026

08:48
Low-cost Custom Fabrication and Mode-locked Operation of an All-normal-dispersion Femtosecond Fiber Laser for Multiphoton Microscopy
Published on: November 22, 2019
8.1K
The truth about laser fiber diameters.
Peter Kronenberg1, Olivier Traxer2
1Urology Department, Hospital Professor Doutor Fernando Fonseca, Amadora, Portugal.
Urology
|November 30, 2014
Summary
Laser fiber diameters vary significantly from advertised values, with outer diameters often much larger. This discrepancy can impact surgical procedures and highlights a lack of manufacturer standardization.
Area of Science:
- Urology
- Medical Device Engineering
- Surgical Technology
Background:
- Laser fibers are critical tools in urological procedures.
- Accurate knowledge of laser fiber dimensions is essential for safe and effective interventions.
- Previous assessments of laser fiber diameter accuracy are limited.
Purpose of the Study:
- To measure actual laser fiber diameters across various manufacturers.
- To compare measured diameters with advertised specifications.
- To identify discrepancies and assess their potential clinical impact.
Main Methods:
- Light microscopy was used to measure outer, cladding, and core diameters of 14 unused laser fibers from 6 manufacturers.
- Advertised diameters ranged from 200 to 400 μm.
- Interviews with industry representatives explored their understanding of advertised diameters.
Main Results:
- Significant differences were found between measured and advertised outer and cladding diameters (P <.00001).
- Outer diameters showed a median increase of 87.3%, with variations up to 116.7%.
- Substantial variations in outer, cladding, and core diameters were observed even for fibers with similar advertised sizes, and labeling lacked clarity.
Conclusions:
- A significant lack of uniformity exists among laser fiber manufacturers regarding diameter specifications.
- Advertised laser fiber diameters are frequently inaccurate, potentially misleading urologists.
- This misinformation can lead to surgical complications due to unexpected fiber sizes impacting interventional parameters.