Percutaneous left atrial appendage closure with a novel self-modelizing device: a pre-clinical feasibility study
Jai-Wun Park1, Mohammad A Sherif1, Konstantin Zintl1
1Department of Cardiology, Angiology, and Pneumology, Coburg Hospital, 96450 Coburg, Germany.
International Journal of Cardiology
|December 3, 2014
Summary
This study shows the Occlutech® left atrial appendage (LAA) occluder is feasible and safe. The device achieved complete LAA closure in all pigs with benign healing and no adverse events, demonstrating its potential for clinical use.
Area of Science:
- Cardiovascular Research
- Medical Device Technology
- Animal Models in Research
Background:
- The left atrial appendage (LAA) is a common source of emboli in conditions like atrial fibrillation.
- Effective LAA occlusion is crucial for preventing stroke.
- New device technologies are needed to improve LAA occlusion safety and efficacy.
Purpose of the Study:
- To evaluate the feasibility and safety of a novel left atrial appendage (LAA) occluder device.
- To assess the device's performance in achieving LAA closure and its long-term integration.
- To determine the biocompatibility and potential complications associated with the LAA occluder.
Main Methods:
- Implantation of the LAA occluder in twelve pigs under TEE and fluoroscopic guidance.
- Assessment of device placement and immediate closure via TEE.
- Macroscopic evaluation of explanted hearts after six weeks to assess device integration and tissue response.
Main Results:
- Successful LAA closure was achieved in 8/10 pigs immediately post-implantation, and in all pigs after six weeks.
- The device demonstrated ease of deployment and excellent adaptation to LAA tissue.
- Macroscopic examination revealed secure device integration, benign healing, and no thrombus formation or significant inflammation. Minor device loop penetration occurred without adverse effects.
Conclusions:
- The Occlutech® LAA occluder is feasible and safe for LAA occlusion.
- The device achieves complete and durable LAA closure with a favorable safety profile.
- The study supports the potential of this device for clinical application in stroke prevention.