Fragility Assessment of Bovine Cortical Bone Using Scratch Tests
Published on: November 30, 2017
Variability in reference point microindentation and recommendations for testing cortical bone: maximum load, sample
T Jenkins1, L V Coutts1, D G Dunlop2
1Bioengineering Science Research Group, Faculty of Engineering and the Environment, University of Southampton, Highfield, Southampton SO17 1BJ, UK.
Reference Point Indentation (RPI) testing protocols were optimized by analyzing key parameters. Consistent application of maximum load, surface preparation, and supported device use are recommended for reliable bone mechanical characterization.
Area of Science:
- Biomechanical Engineering
- Materials Science
- Orthopedic Research
Background:
- Reference Point Indentation (RPI) is a promising technique for assessing bone fracture risk and mechanical properties.
- Standardized testing protocols are lacking, hindering consistent application of RPI in clinical and laboratory settings.
Purpose of the Study:
- To investigate the influence of key testing parameters on the Indentation Distance Increase (IDI) measured by RPI.
- To provide evidence-based recommendations for optimizing RPI test protocols for bone analysis.
Main Methods:
- The RPI tool was used to indent bovine and human bone samples (femora, femoral heads, and necks).
- Five parameters were evaluated: maximum load, sample orientation, mode of use (freehand vs. supported), surface preparation, and measurement spacing.
- The Indentation Distance Increase (IDI) was analyzed in relation to each parameter.
Main Results:
- IDI is significantly affected by maximum load, mode of use, and surface preparation, necessitating consistency in these factors.
- Supported device use and bone surface preparation (periosteum removal, machining) reduce the coefficient of variation.
- A sample thickness at least 10 times the indentation depth is recommended for representative measurements.
- Measurement spacing did not significantly influence IDI, with 500 µm being adequate.
Conclusions:
- Optimizing RPI parameters, including consistent load, supported use, and appropriate surface preparation, enhances measurement reliability.
- Adhering to recommended protocols minimizes variability and improves the accuracy of bone mechanical characterization using RPI.
- These findings support the broader clinical and laboratory application of RPI for bone assessment.