Quantitative comparison between the commercial software STRATOS(®) by Philips and a homemade software for
Background:
Targeted radionuclide therapy is a rapidly growing modality. A few commercial treatment planning systems are entering the market. However, some in-house systems are currently developed for a more flexible and customized dosimetry calculation at voxel-level. For this purpose, we developed a novel software, VoxelMed, and performed a comparison with the software STRATOS.
Methods:
The validation of both of them was undertaken using radioactive phantoms with different volume inserts. A cohort of 10 patients was also studied after a therapeutic administration of (177)Lu-labelled radiopeptides. The activity, number of disintegrations, absorbed dose and dose-volume histogram (DVH) were calculated for the phantoms and the kidneys in patients, which were the main critical organs at risk in this study.
Results:
In phantoms the absorbed doses computed with VoxelMed and STRATOS agree within 5%. In patients at the voxel-level the absorbed dose to kidneys (VoxelMed: mean 0.66 Gy/GBq) showed a limited difference of 5%, but with a remarkable range (-40%, +60%) between the two software packages. Voxel-dosimetry allows to estimate the dose non-homogeneities in volumes, which may be evaluated through DVHs.
Conclusion:
This study demonstrates that a fully 3D voxel-dosimetry with multiple SPECT images is feasible by using home-made or commercial software package and absorbed dose results obtained are similar. The main difference between the studied tools was observed in the activity integration method (effective vs physical half-time to time activity curve tail). We believe that an effective half-time integration method produces a more accurate approximation of clinical uptake and resultant dosimetry.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET