Feasibility of ultra-low radiation dose reduction for renal stone CT using model-based iterative reconstruction:
J Scott Kriegshauser1, Sailen G Naidu1, Robert G Paden1
1Department of Radiology, Mayo Clinic Hospital, Phoenix, AZ.
Purpose:
To compare dose, reader assessments, and noise between ultra-low-dose (ULD) and low-dose (LD) renal stone computed tomography (CT) using multiple reconstruction methods.
Materials And Methods:
Prospective study of 19 patients having noncontrast LD and ULD renal stone CT reconstructed using filtered back projection, adaptive statistical iterative reconstruction (ASIR), and model-based iterative reconstruction (MBIR). Dose, noise measurements, and subjective image assessments were recorded.
Results:
ULD volume CT dose index was 61% less. Noise was 50% less with ULD MBIR than with LD ASIR (P<.001); no other significant differences existed.
Conclusion:
ULD MBIR is comparable to our standard of care, LD ASIR, with 61% lower dose and 50% less noise.