[Chamber Angle Assessment in Clinical Practice - A Comparison between Optical Coherence Tomography and Gonioscopy]
M P Mösler1, J U Werner1, G K Lang1
1Augenklinik, Universitätsklinikum Ulm.
Summary
Anterior segment optical coherence tomography (AS-OCT) is an effective alternative to gonioscopy for assessing glaucoma chamber angles, especially when gonioscopy is difficult. AS-OCT, combined with anterior chamber depth measurement, provides valuable prognostic information for angle closure risk.
Area of Science:
- Ophthalmology
- Medical Imaging
- Glaucoma Research
Background:
- Anterior chamber structures are crucial for glaucoma classification, therapy, progression, and prognosis.
- Anterior segment optical coherence tomography (AS-OCT) is increasingly relevant for evaluating these structures.
- This study compares the diagnostic performance of AS-OCT with gonioscopy for chamber angle assessment in glaucoma patients.
Purpose of the Study:
- To compare the diagnostic accuracy of AS-OCT and gonioscopy in assessing glaucoma chamber angles.
- To evaluate anterior chamber depth (ACD) measured by slit-lamp examination as a prognostic factor for chamber angle width (CAW).
- To determine the clinical utility of AS-OCT in routine glaucoma management.
Main Methods:
- 104 glaucoma patients underwent AS-OCT imaging.
- Chamber angle grading from AS-OCT was compared to gonioscopic grading (Shaffer system).
- Anterior chamber depth (ACD) was measured using slit-lamp examination and AS-OCT to assess its correlation with chamber angle width (CAW).
Main Results:
- AS-OCT revealed discrepancies with gonioscopy in 17% of wide angles and 35% of narrow angles.
- Anterior chamber depth (ACD) assessment is a strong predictor of CAW (correlation coefficient 0.83), with a critical ACD of 2.4 mm for angle closure risk.
- A 1 mm variation in ACD correlated with an 18.9° change in CAW; all angle closure glaucoma patients had ACD < 2.4 mm.
Conclusions:
- AS-OCT is an effective alternative to gonioscopy for chamber angle assessment in glaucoma.
- AS-OCT can be superior to gonioscopy in certain clinical situations, particularly with inexperienced practitioners or when gonioscopy is challenging.
- ACD measurement via AS-OCT provides valuable prognostic information for angle closure risk, complementing gonioscopy findings.
More Related Videos
Related Concept Videos
Angle Closure Glaucoma: Treatment
1.7K
Angle-closure glaucoma, or closed-angle glaucoma, is an eye condition where the iris bulges out and blocks the iridocorneal angle, resulting in a buildup of aqueous humor and increased intraocular pressure. Immediate medical attention is necessary due to the sudden onset of symptoms. The treatment for angle-closure glaucoma includes short-term and long-term approaches. Short-term treatment involves using eye drops like pilocarpine to lower intraocular pressure by increasing aqueous humor...
1.7K
Open Angle Glaucoma: Treatment
1.1K
In open-angle glaucoma, the iridocorneal angle remains open, but the trabecular meshwork becomes stiff, slowing down the outflow of aqueous humor. This causes a buildup of aqueous humor in the anterior chamber, leading to a sudden increase in intraocular pressure. The treatment for open-angle glaucoma focuses on reducing the elevated intraocular pressure by either decreasing the secretion of aqueous humor or increasing its outflow.
Drugs such as carbonic anhydrase inhibitors, α2- and...
Drugs such as carbonic anhydrase inhibitors, α2- and...
1.1K
Glaucoma: Overview
1.7K
Glaucoma is an eye condition characterized by increased intraocular pressure that damages the retina and optic nerve, leading to irreversible blindness if left untreated. The human eye has various components, including the cornea, iris, pupil, lens, and optic nerve. Aqueous humor is secreted by the epithelium of the ciliary body in the posterior chamber and flows through the trabecular meshwork and canal of Schlemm, maintaining normal intraocular pressure. The trabecular meshwork and the canal...
1.7K