Related Experiment Video
Updated: Apr 20, 2026

Non-invasive Optical Measurement of Cerebral Metabolism and Hemodynamics in Infants
Published on: March 14, 2013
Non-invasive measurement of brain temperature using radiometric thermometry: experimental validation and clinical
W Thomas Bass1, F A Lattanzio2, G Brayman1
1Department of Pediatrics, Divisions of Neonatal Medicine, Eastern Virginia Medical School, Children's Hospital of The King's Daughters, Norfolk, VA, USA.
Background:
Therapeutic hypothermia (HT) has been shown to decrease death and severe disability in infants with hypoxic-ischemic encephalopathy (HIE). Rectal temperature (RT) is used to determine the temperature set-points for treatment with HT, however experimental studies have shown significant differences between RT and brain temperature during HT. Knowledge of actual brain temperature during HT might allow better determination of optimal degree of cooling and improve outcomes.
Objectives:
To compare measurements of brain temperature obtained by non-invasive radiometric thermometry (RadT) to direct tissue measurements in an experimental model of HT, and to use RadT in newborn infants with HIE undergoing HT.
Study Design:
RadT measurements of brain temperature were compared to fiber optic (Luxtron) thermometry measurements placed at a depth of 1.5 centimeters into the brain of cooled miniswine. Following validation studies, brain RadT and RT measurements were continuously recorded in thirty infants with HIE during HT and rewarming.
Results:
RadT and Luxtron probe temperatures were comparable in miniswine throughout a temperature range similar to therapeutic HT. RadT measurements of brain temperature were higher than RT in 60% of infants with HIE undergoing HT. Higher RadT measurements compared to RT were associated with cerebral white matter abnormalities (p = 0.01).
Conclusions:
RadT provides a safe, passive and non-invasive way to measure brain temperature that can be used in the clinical setting. RadT may be helpful in determining the optimal degree of cooling and identifying infants at highest risk of brain injury.
More Related Videos
07:27How to Administer Near-Infrared Spectroscopy in Critically ill Neonates, Infants, and Children
Published on: August 19, 2020
06:29Electrophysiological Measurement of Noxious-evoked Brain Activity in Neonates Using a Flat-tip Probe Coupled to Electroencephalography
Published on: November 29, 2017
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...