Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Computed Tomography01:10

Computed Tomography

9.7K
Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
9.7K
Imaging Studies III: Computed Tomography01:27

Imaging Studies III: Computed Tomography

828
DefinitionComputed Tomography (CT) of the genitourinary (GU) tract is a non-invasive imaging modality that utilizes X-rays and computer processing to generate detailed cross-sectional images of the urinary system, encompassing the kidneys, ureters, bladder, and adjacent structures such as the adrenal glands.PurposeCT scans of the GU tract serve several diagnostic and therapeutic purposes, including:Diagnosis of Urinary Tract Diseases: Detects kidney stones, tumors, cysts, and congenital...
828
Electron Microscope Tomography and Single-particle Reconstruction01:07

Electron Microscope Tomography and Single-particle Reconstruction

3.1K
Transmission electron microscopy (TEM) can be used to determine the 3D structure of biological samples with the help of techniques such as electron microscope tomography and single-particle reconstruction. While single-particle reconstruction can examine macromolecules and macromolecular complexes in vitro conditions only, tomography permits the study of cell components or small cells in vivo.
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
3.1K
Imaging Studies I: CT and MRI01:14

Imaging Studies I: CT and MRI

1.3K
Introduction: MRI and CT scans are crucial advancements in medical imaging techniques, playing a vital role in diagnosing conditions related to the gastrointestinal (GI) system. Each scan serves distinct purposes, targets specific areas, and requires unique nursing duties.
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
1.3K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Efficacy and Safety of Thalidomidein Patients with Transfusion-Dependent B-Thalassemia: A Systematic Review, Meta-Analysis and GRADE Evaluation.

Mediterranean journal of hematology and infectious diseases·2026
Same author

Proton therapy range uncertainty reduction using vendor-agnostic tissue characterization on a virtual photon-counting CT head scan.

Medical physics·2026
Same author

Optimizing energy settings in CdTe, CZT, and Si photon-counting CT for material separation and detection.

Physics in medicine and biology·2026
Same author

Precise Lung Density Quantification with a Physics-based CT Harmonizer.

Radiology. Cardiothoracic imaging·2026
Same author

Diagnostic Value of Cerebrospinal Fluid Metagenomics Next-generation Sequencing in Neurobrucellosis in Children: Erratum.

The Pediatric infectious disease journal·2026
Same author

Demographic distribution matching between real-world and virtual phantom population.

Medical physics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
Same journal

A novel optical respiratory gating system with a hybrid phase-amplitude algorithm for spot-scanning proton therapy.

Medical physics·2026
Same journal

Gamma Knife treatment planning using knowledge-based reinforcement learning.

Medical physics·2026
Same journal

Development and characterization of a novel, small animal external beam irradiator using a clinical high dose rate brachytherapy source.

Medical physics·2026
Same journal

Deep learning-based dose prediction for MR-guided prostate SIB: Supporting rapid feasibility assessment and adaptive editing margin selection.

Medical physics·2026
See all related articles

Related Experiment Video

Updated: Apr 20, 2026

Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
12:32

Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans

Published on: September 27, 2020

10.4K

Evaluating iterative reconstruction performance in computed tomography.

Baiyu Chen1, Juan Carlos Ramirez Giraldo2, Justin Solomon1

  • 1Medical Physics Graduate Program, Duke University, Durham, North Carolina 27705 and Carl E. Ravin Advanced Imaging Laboratories, Duke University, Durham, North Carolina 27705.

Medical Physics
|December 5, 2014
PubMed
Summary
This summary is machine-generated.

Iterative reconstruction (IR) in computed tomography (CT) significantly reduces radiation dose compared to filtered backprojection reconstruction (FBP). IR algorithms offer superior performance across various tasks, especially for challenging low-contrast, small-size targets.

More Related Videos

Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
05:32

Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph

Published on: February 21, 2025

796
High Resolution 3D Imaging of Ex-Vivo Biological Samples by Micro CT
08:57

High Resolution 3D Imaging of Ex-Vivo Biological Samples by Micro CT

Published on: June 21, 2011

19.4K

Related Experiment Videos

Last Updated: Apr 20, 2026

Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
12:32

Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans

Published on: September 27, 2020

10.4K
Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
05:32

Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph

Published on: February 21, 2025

796
High Resolution 3D Imaging of Ex-Vivo Biological Samples by Micro CT
08:57

High Resolution 3D Imaging of Ex-Vivo Biological Samples by Micro CT

Published on: June 21, 2011

19.4K

Area of Science:

  • Medical Imaging Physics
  • Radiological Sciences
  • Computational Imaging

Background:

  • Iterative reconstruction (IR) offers advantages in computed tomography (CT) but its nonlinear behavior complicates performance characterization.
  • Traditional assessments of IR performance often rely on contrast and noise metrics, potentially overlooking task-specific detectability.

Purpose of the Study:

  • To evaluate the performance of IR in CT using both task-specific and task-generic strategies.
  • To quantify the dose reduction potential of different IR algorithms compared to filtered backprojection reconstruction (FBP).

Main Methods:

  • Mathematical assessment of IR performance using an observer model to predict detection accuracy (d') based on image noise, resolution, and task parameters.
  • Modeling of various tasks (1-4 mm sizes, 10-100 HU contrast, sharp/soft edges) across five radiation dose levels (CTDIvol: 3.4-64.8 mGy).
  • Calculation of threshold dose for achieving a threshold AUC of 0.9, and derivation of task-specific dose reduction potential and task-generic performance comparisons.

Main Results:

  • IR algorithms required less radiation dose than FBP to achieve diagnostic image quality (threshold AUC of 0.9).
  • Significant dose reduction potentials were observed, with SAFIRE5 (77%-84%), SAFIRE3 (50%-61%), and IRIS (37%-50%), particularly for challenging tasks (low contrast, small size).
  • Task-generic comparisons showed IR's overall superiority, with IRIS (8%-12%), SAFIRE3 (10%-16%), and SAFIRE5 (20%-33%) enabling a larger percentage of tasks to exceed the performance threshold.

Conclusions:

  • Task-specific evaluations reveal that IR's dose reduction potential is dependent on task attributes, enabling optimization for specific clinical indications.
  • Task-generic evaluations confirm IR's overall superiority to FBP, providing a basis for general algorithm comparisons.
  • IR demonstrates significant potential for dose reduction and improved performance in CT imaging across a wide range of clinical tasks.