Evaluation of Drug Sorption to PVC- and Non-PVC-based Tubes in Administration Sets Using a Pump
Published on: March 11, 2017
An in vitro evaluation of infusion methods using a syringe pump to improve noradrenaline administration
S Genay1, B Décaudin, S Scoccia
1Department of Biopharmacy, Galenic and Hospital Pharmacy, UDSL, EA GRIIOT, UFR Pharmacie, Lille 2 University, Lille, France; Pharmacy, Lille University Hospital, Lille, France.
Background:
International guidelines recommend noradrenaline (NA) as the vasopressor of choice to treat septic shock. The aim of this study was to determine the best way to infuse patients with NA.
Methods:
The in vitro study was designed to measure NA concentration at the end of each studied assembly line. Three infusion systems used the double pump method and three single pumps, which differed as regards NA concentrations (0,2 - 0,5 - 1 mg/h), dead space volume of the devices and the use of saline. Infusion systems were compared according to the time necessary to reach an NA mass flow rate steady-state plateau after the onset of infusion or after a flow change.
Results:
Times were significantly different between the six methods for infusing NA. The system using the double syringe method with a standard extension set was the longest to reach the steady state after the onset of infusion [40.00 min (19.57 - 49.22)]. The steady-state plateau was obtained most rapidly with the double-syringe pump systems using very low dead-space volume extension sets and single-syringe pump systems containing diluted noradrenaline at the beginning of NA infusion.
Conclusion:
A combination of a low dead-space volume extension set and a double pump method with a constant saline flow rate at 5 ml/h might be the solution to provide the most reliable NA infusion delivery.
Related Concept Videos
IV Infusion to Oral Dosing: Conversion Methods
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
In Vitro Drug Dissolution: Compendial Testing Models II
Routes of Drug Administration: Parenteral
The intravenous route (IV) of drug administration can be further categorized into two types. The bolus injection administers the entire dose rapidly, while an intravenous infusion slowly delivers smaller doses steadily.
The IV route is often...