Finger volume pulse waveforms facilitate reliable assessment of heart rate variability, but not blood pressure
Jonathan R Linder, Harald M Stauss, Holly Gindes
1Department of Internal Medicine, The University of Iowa, Iowa City, IA 52242, USA. jess-fiedorowicz@uiowa.edu.
Background:
We sought to determine whether heart rate variability (HRV), blood pressure (BP) variability, and baroreceptor-heart rate reflex sensitivity can be reliably assessed using finger volume pulse waveforms obtained from the commercially available EndoPAT device.
Methods:
Non-invasive BP (Finometer Pro as a non-invasive standard) and finger volume (EndoPAT) waveforms were recorded in 65 adults (37 ± 14 years; 60% female) and systolic BP and heart rate (HR) time series were derived after calibrating the EndoPAT signal based on systolic and diastolic BP values obtained by a sphygomomanometer. Transfer function analyses were performed to test for coherence between systolic BP and HR time series derived from the Finometer and EndoPAT devices. Time-domain HRV parameters, frequency domain HR and systolic BP variability parameters, and baroreflex sensitivity (sequence technique) were computed from Finometer- and EndoPAT-derived time series and intraclass correlation coefficients (ICC) were calculated.
Results:
Squared coherence between systolic BP time series derived from the Finometer and EndoPAT devices was low, suggesting poor correlation. In contrast, squared coherence between HR time series derived from the two devices was excellent [High Frequency (HF) = 0.80, Low Frequency (LF) = 0.81], with gain values close to 1.0. ICC values for time- and frequency-domain HRV parameters were excellent (>0.9 except for relative HF HRV, which was 0.77), while ICC values for frequency-domain BP variability parameters and baroreceptor-HR reflex sensitivity were low.
Conclusions:
Finger volume pulse waveforms can be used to reliably assess both time-domain and frequency-domain HR variability. However, frequency domain BP variability parameters cannot be reliably assessed from finger volume pulse waveforms using the simple calibration technique used in this study.
More Related Videos
Related Concept Videos
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Special considerations while measuring pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Regulation of Pulse
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: