Accuracy and user performance evaluation of a blood glucose monitoring system
Michael Caswell1, Joy Frank, Maria T Viggiani
11 Consumer Product Testing Company, Inc. , Fairfield, New Jersey.
Background:
The accuracy of the Contour(®) Plus (Bayer HealthCare LLC, Diabetes Care, Whippany, NJ) blood glucose monitoring system (BGMS) was evaluated in two separate studies.
Materials And Methods:
In the laboratory study, fingerstick samples from 100 subjects were tested in duplicate using three test strip lots and assessed per International Organization for Standardization (ISO) 15197:2003, Section 7 (≥95% of results within ±15 mg/dL or ±20% of the reference result for samples with glucose concentrations <75 and ≥75 mg/dL, respectively) and ISO 15197:2013, Section 6.3 (≥95% of results within ±15 mg/dL or ±15% of the reference result for samples with glucose concentrations <100 and ≥100 mg/dL, respectively) accuracy criteria. In the clinical trial, 220 subjects with diabetes, naive to the BGMS, tested capillary glucose from fingertip and palm blood samples and completed an ease-of-use questionnaire. BGMS and YSI glucose analyzer results were compared.
Results:
In the laboratory study, 100% of results met ISO 15197:2003 and ISO 15197:2013 accuracy criteria. In the clinical trial, 100% and 99.1% of subject fingerstick results and 98.1% and 96.7% of subject palm results met ISO 15197:2003 and ISO 15197:2013 accuracy criteria, respectively. By Parkes Consensus Error Grid analysis, 100% of subject fingerstick results and 98.1% of subject palm results were within Zone A (remainder within Zone B). Questionnaire results showed most subjects found the BGMS easy to use.
Conclusions:
The Contour Plus BGMS meets ISO 15197:2003 and ISO 15197:2013 accuracy criteria in the laboratory and when used by untrained individuals.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...