The effectiveness of the Spot Vision Screener in detecting amblyopia risk factors
Mae Millicent W Peterseim1, Carrie E Papa1, M Edward Wilson1
1Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina.
Purpose:
To evaluate the updated Spot Vision Screener (PediaVision, Welch Allyn, Skaneateles Falls, NY) in detecting amblyopia risk factors using 2013 guidelines of American Association for Pediatric Ophthalmology and Strabismus (AAPOS).
Methods:
In this prospective study, patients seen from June 2012 to November 2013 were tested with the Spot prior to examination by a pediatric ophthalmologist who was masked to test results. The following data were analyzed: age, subject testability, examination findings, and systemic and ocular pathology. Children were divided into three age groups to determine gold standard results according to the AAPOS guidelines.
Results:
A total of 444 children (average age, 72 months) were included. Compared to the ophthalmologist's examination, the Spot sensitivity was 87.7% and the specificity was 75.9% in detecting amblyopia risk factors. Sensitivity did not differ significantly between age groups, although the positive predictive value improved in the older age groups.
Conclusions:
In our study cohort, the Spot provided good specificity and sensitivity in detecting amblyopia risk factors according 2013 AAPOS criteria, with minor improvements with updated versions.