A model-based analysis of a simplified beam-specific dose output in proton therapy with a single-ring wobbling system
Yuki Kase1, Haruo Yamashita, Masumi Numano
1Proton Therapy Division, Shizuoka Cancer Center Hospital, 1007, Shimonagakubo, Nagaizumi-cho, Shizuoka 411-8777, Japan. Proton Therapy Division, Shizuoka Cancer Center Research Institute, 1007, Shimonagakubo, Nagaizumi-cho, Shizuoka 411-8777, Japan.
Abstract:
In radiation therapy, it is necessary to preset a monitor unit in an irradiation control system to deliver a prescribed absolute dose to a reference point in the planning target volume. The purpose of this study was to develop a model-based monitor unit calculation method for proton-beam therapy with a single-ring wobbling system. The absorbed dose at a calibration point per monitor unit had been measured for each beam-specific measurement condition without a patient-specific collimator or range compensator before proton therapeutic irradiation at Shizuoka Cancer Center. In this paper, we propose a simplified dose output model to obtain the output ratio between a beam-specific dose and a reference field dose, from which a monitor unit for the proton treatment could be derived without beam-specific measurements. The model parameters were determined to fit some typical data measured in a proton treatment room, called a Gantry 1 course. Then, the model calculation was compared with 5456 dose output ratios that had been measured for 150-, 190- and 220 MeV therapeutic proton beams in two treatment rooms over the past decade. The mean value and standard deviation of the difference between the measurement and the model calculation were respectively 0.00% and 0.27% for the Gantry 1 course, and -0.25% and 0.35% for the Gantry 2 course. The model calculation was in good agreement with the measured beam-specific doses, within 1%, except for conditions less frequently used for treatment. The small variation for the various beam conditions shows the high long-term reproducibility of the measurement and high degree of compatibility of the two treatment rooms. Therefore, the model was expected to assure the setting value of the dose monitor for treatment, to save the effort required for beam-specific measurement, and to predict the dose output for new beam conditions in the future.
More Related Videos
Related Concept Videos
Singularity Functions for Bending Moment
Beams with Unsymmetric Loadings
The first moment-area theorem determines the slope at any point on the beam. This theorem indicates that the change in slope between two points on a beam...
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...
Elastic Curve from the Load Distribution
For all beams, the analysis of the beam's reaction to distributed loads begins by understanding the relationship between a beam's load and the resulting shear forces and bending moments. Initially, this...
Beams with Symmetric Loadings
The M/EI...
Design of Prismatic Beams for Bending