Improvement of the trigger of a ventilator for non-invasive ventilation in children: bench and clinical study
Related Experiment Video
Updated: Apr 19, 2026

Inspiratory Muscle Training as an Adjunct to the Treatment of Weaning Failure in Critically Ill Patients: A Practical Guide
Published on: January 30, 2026
Sonia Khirani1,2, Bruno Louis3, Karl Leroux4
1S2A Santé, Ivry-sur-Seine, France.
Insights
An improved inspiratory trigger (IT) for ventilators enhances synchronization with pediatric respiratory efforts. This novel approach, tested on a bench and in young patients, significantly reduces trigger time delays and pressures for better non-invasive ventilation (NIV) support.
Area of Science:
- Pediatric respiratory support
- Mechanical ventilation
- Biomedical engineering
Background:
- Limited availability of ventilators specifically designed for pediatric use.
- Existing home ventilators may exhibit poor synchronization with children's respiratory efforts.
- Inappropriate inspiratory trigger (IT) settings in assist modes for pediatric patients.
Purpose of the Study:
- To evaluate the efficacy of an improved inspiratory trigger (IT) for ventilators in pediatric applications.
- To assess the synchronization capabilities of the enhanced IT on a pediatric test lung and in young patients.
Main Methods:
- Comparison of a classical IT (ITc) with an improved IT incorporating non-invasive ventilation (NIV) + IT.
- Testing conducted on a lung model simulating six pediatric breathing profiles.
- Clinical evaluation involving six pediatric patients (mean age 14.1 years) requiring long-term NIV.
Main Results:
- Bench testing demonstrated significant reductions in trigger time delays (ΔT) and trigger pressures (ΔP) with the NIV + IT compared to ITc.
- Clinical study confirmed a significant decrease in ΔT and ΔP in pediatric patients using the improved IT.
- The improved IT showed enhanced sensitivity and responsiveness for pediatric non-invasive ventilation.
Conclusions:
- The sensitivity of ventilator inspiratory triggers can be effectively improved for pediatric use.
- The enhanced IT (NIV + IT) offers superior performance compared to conventional ITs in pediatric patients.
- Findings support the clinical utility of the improved IT for pediatric non-invasive ventilation.
Background And Aims:
Even though numerous ventilators are licensed for a use in children, very few have been specifically developed for this age range. Therefore, home ventilators may not be able to adequately synchronize with the child's respiratory effort, and the inspiratory triggers (ITs) of assist modes are not always appropriate for children. The aim of the study was to test the improvement of the IT of a ventilator on a pediatric bench and in pediatric patients.
Methods:
A classical IT (ITc) and an improved IT [non-invasive ventilation (NIV) + IT] were tested on a bench with six pediatric profiles and in six young patients (mean age 14.1 ± 2.7 years old) requiring long-term NIV.
Results:
On the bench, trigger time delays (ΔT) and trigger pressures (ΔP) were reduced with the NIV + IT as compared with the ITc (ΔT: 0.481 ± 0.332 vs 0.079 ± 0.022 s for ITc and NIV + IT, respectively, P = 0.027; ΔP: -1.40 ± 0.70 vs -0.42 ± 0.28 cmH2 O for ITc and NIV + IT, respectively, P = 0.046). The clinical study confirmed the decrease in ΔT (0.267 ± 0.061 vs 0.178 ± 0.074 s for ITc and NIV + IT, respectively, P = 0.024) and ΔP (-0.68 ± 0.26 vs -0.39 ± 0.11 cmH2 O for ITc and NIV + IT, respectively, P = 0.030).
Conclusions:
The sensitivity of the IT of a ventilator can be improved for pediatric use. The improvements observed on the bench study were confirmed in pediatric patients.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...