Study of ageing and fading in CR-39 detectors for different storage conditions
Daniele Franci1, Tommaso Aureli2, Francesco Cardellini3
1Department of Rome, ARPA Lazio, Via Giuseppe Saredo 52, Rome 00100, Italy daniele.franci@arpalazio.it.
Abstract:
The aim of this study was to investigate the effect of ageing and fading on PADC detector response, as a function of the storage time and temperature. Several groups of CR-39 detectors provided by Radosys, Ltd. were exposed at the reference radon chamber of the Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, at the ENEA centre in Casaccia. The results indicate that low-temperature storage inhibits the effect of both ageing and fading. Finally, the overall reduction in CR-39 sensitivity due to the combined ageing/fading effect was estimated. In particular, the sensitivity of the detectors continuously exposed in air at room temperature over 6 and 3 months was reduced, respectively, by 7.5 and 4 %.