Automated analysis of flow cytometric data for measuring neutrophil CD64 expression using a multi-instrument
Linda Wong1, Beth L Hill2, Benjamin C Hunsberger2
1Trillium Diagnostics, LLC, Brewer, Maine.
Cytometry. Part B, Clinical Cytometry
|December 23, 2014
Summary
A new automated method using GemStone and probability state modeling (PSM) accurately measures neutrophil CD64 expression, improving infection and inflammation diagnostics. This approach significantly reduces interanalyst variability compared to the current semi-automated Leuko64 assay.
Area of Science:
- Flow cytometry
- Immunology
- Infectious disease diagnostics
Background:
- Leuko64™ assay measures neutrophil CD64 expression as an indicator of infection, sepsis, or systemic inflammation.
- Current Leuko64 analysis uses QuantiCALC software, a semi-automated method with limited interanalyst variability (<5%).
- This study introduces and evaluates a fully automated approach using GemStone™ software and probability state modeling (PSM).
Purpose of the Study:
- To evaluate a fully automated method for measuring neutrophil CD64 expression using GemStone and PSM.
- To compare the performance of the automated method against the existing semi-automated QuantiCALC method.
- To assess the impact of automation on interanalyst variability and precision in CD64 index quantification.
Main Methods:
- Processed 457 human blood samples using the Leuko64 assay across four different flow cytometer models.
- Developed a PSM to automatically identify calibration beads and leukocyte subpopulations.
- PSM automatically calculated CD64 index values, with data analysis and quality control performed by GemStone software without operator intervention.
Main Results:
- The automated PSM with GemStone showed excellent correlation with expert manual analysis (r² = 0.99675) for neutrophil CD64 index values.
- Interanalyst imprecision was reduced from 1.06% with QuantiCALC to 0.00% with GemStone PSM.
- Operator-to-operator agreement in GemStone was a perfect correlation (r² = 1.000).
Conclusions:
- Automated quantification of CD64 index values using PSM strongly correlates with expert analysis.
- PSM successfully processed flow cytometric data from multiple instruments and kit lots.
- The probability-based method offers enhanced objectivity, speed, and precision for in vitro diagnostic flow cytometry assays.