Effect of secondary particles on image quality of dynamic flat panels in carbon ion scanning beam treatment
11 Research Center for Charged Particle Therapy, Medical Physics Research Program, National Institute of Radiological Sciences, Chiba, Japan.
Objective:
Real-time markerless tumour tracking using radiographic fluoroscopic imaging is one of the better solutions to improving respiratory-gated radiotherapy. However, particle beams cause secondary particles from patients, which could affect radiographs. Here, we evaluated the quality of radiographs during carbon ion pencil beam scanning (CPBS) irradiation for respiratory gating.
Methods:
A water phantom and chest phantom were used. The phantoms were irradiated with CPBS at 290 MeV n(-1) from orthogonal directions. Dose rates were 3.4 × 10(8), 1.14 × 10(8) and 3.79 × 10(7) particles per second. A dynamic flat panel detector (DFPD) was installed on the upstream (DFPD1) or downstream (DFPD2) side of the vertical irradiation port. DFPD images were acquired during CPBS at 15.00, 7.50 and 3.75 frames per second (fps). Charge on the DFPD was cleaned using fast readout technique every 30 fps. DFPD images were acquired during CPBS with radiographic exposure, and results with and without fast readout technique were compared.
Results:
Secondary particles were visualized as spots or streak-like shapes. Capture of secondary particles from the horizontal beam direction was lower with fast readout technique than without it. With regard to beam irradiation direction dependency, CPBS from the horizontal direction resulted in a greater magnitude of secondary particles reaching DFPD2 than reaching DFPD1. When CPBS was delivered from the vertical direction, however, the magnitude of secondary particles on both DFPDs was very similar.
Conclusion:
Fast readout technique minimized the effect of secondary particles on DFPD images during CPBS.
Advances In Knowledge:
This technique may be useful for markerless tumour tracking for respiratory gating.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
10:18Dynamic Pore-scale Reservoir-condition Imaging of Reaction in Carbonates Using Synchrotron Fast Tomography
Published on: February 21, 2017
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...