Quantitative Micro-CT Analysis of Aortopathy in a Mouse Model of β-aminopropionitrile-induced Aortic Aneurysm and Dissection
Published on: July 16, 2018
Automatic characterization of thoracic aortic aneurysms from CT images
Juan Antonio Martínez-Mera1, Pablo G Tahoces1, José M Carreira2
1Centro Singular de Investigación en Tecnoloxías da Información (CITIUS), University of Santiago de Compostela, Spain.
An automated method accurately measures thoracic aorta diameter using principal component analysis. This technique shows excellent agreement with gold standards and robust rotational invariance, aiding in aneurysm diagnosis.
Area of Science:
- Medical Imaging
- Cardiovascular Imaging
- Image Analysis
Background:
- Accurate thoracic aorta diameter measurement is crucial for diagnosing and monitoring aortic abnormalities like aneurysms.
- Existing methods may lack automation or robustness in diverse clinical scenarios.
Purpose of the Study:
- To develop and validate an automatic method for measuring thoracic aorta diameter.
- To assess the accuracy and rotational invariance of the proposed measurement technique.
Main Methods:
- Principal Component Analysis (PCA) applied to normal planes of the aorta to define its main axis.
- Validation using a database of 15 clinical cases compared against commercial software (gold standard).
- Rotational invariance tested using phantom images in various orientations.
Main Results:
- Good agreement observed between the automated method and the gold standard in clinical cases, with a bias of 0.2mm.
- Bland-Altman plots confirmed values within acceptable limits of agreement.
- ANOVA test showed no statistically significant difference in results for phantom images at different orientations (F=1.88, p=0.153), demonstrating robustness.
Conclusions:
- The developed automatic method provides accurate and reliable thoracic aorta diameter measurements.
- The technique exhibits robustness to rotational variations, making it suitable for clinical application.
- This approach can aid in the automatic characterization of thoracic aortic aneurysms.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:08Ultrasound Imaging of the Thoracic and Abdominal Aorta in Mice to Determine Aneurysm Dimensions
Published on: March 8, 2019
Related Concept Videos
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Imaging Studies for Cardiovascular System V: CT
Aneurysm I: Introduction
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Aortic Regurgitation II: Clinical Features and Diagnostic Tests