Pitfalls and premature failure of the Freedom SOLO stentless valve
Olaf Stanger1, Irina Bleuel2, Sylvia Reineke2
1Clinic for Cardiovascular Surgery, Inselspital, Bern University Hospital and University of Bern, Bern, Switzerland olaf.stanger@insel.ch.
Objectives:
This study reports a series of pitfalls, premature failures and explantation of the third-generation Freedom SOLO (FS) bovine pericardial stentless valve.
Methods:
A total of 149 patients underwent aortic valve replacement using the FS. Follow-up was 100% complete with an average observation time of 5.5 ± 2.3 years (maximum 8.7 years) and a total of 825 patient-years. Following intraoperative documentation, all explanted valve prostheses underwent histological examination.
Results:
Freedom from structural valve deterioration (SVD) at 5, 6, 7, 8 and 9 years was 92, 88, 80, 70 and 62%, respectively. Fourteen prostheses required explantation due to valve-independent dysfunction (n = 5; i.e. thrombus formation, oversizing, aortic dilatation, endocarditis and suture dehiscence) or valve-dependent failure (acute leaflet tears, n = 4 and severe stenosis, n = 5). Thus, freedom from explantation at 5, 6, 7, 8 and 9 years was 95, 94, 91, 81 and 72%, respectively. An acute vertical tear along the non-coronary/right coronary commissure to the base occurred at a mean of 6.0 years (range 4.3-7.3 years) and affected size 25 and 27 prostheses exclusively. Four FS required explantation after a mean of 7.5 years (range 7.0-8.3 years) due to severe functional stenosis and gross calcification that included the entire aortic root.
Conclusions:
The FS stentless valve is safe to implant and shows satisfying mid-term results in our single institution experience. Freedom from SVD and explantation decreased markedly after only 6-7 years, so that patients with FS require close observation and follow-up. Exact sizing, symmetric positioning and observing patient limitations are crucial for optimal outcome.
More Related Videos
14:14Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Related Concept Videos
Mitral Stenosis I: Introduction
Aortic Regurgitation I: Introduction
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Valve Prolapse I: Introduction
Mitral Stenosis III: Medical Management
Aortic Regurgitation II: Clinical Features and Diagnostic Tests