Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental
Yasser Morera-Gómez1, Héctor A Cartas-Aguila1, Carlos M Alonso-Hernández1
1Centro de Estudios Ambientales de Cienfuegos, AP 5. Ciudad Nuclear, CP 59350 Cienfuegos, Cuba.
Abstract:
Monte Carlo efficiency transfer method was used to determine the full energy peak efficiency of a coaxial n-type HPGe detector. The efficiencies calibration curves for three Certificate Reference Materials were determined by efficiency transfer using a (152)Eu reference source. The efficiency values obtained after efficiency transfer were used to calculate the activity concentration of the radionuclides detected in the three materials, which were measured in a low-background gamma spectrometry system. Reported and calculated activity concentration show a good agreement with mean deviations of 5%, which is satisfactory for environmental samples measurement.
More Related Videos
Related Concept Videos
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...