Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Interfractional variations of tumor centroid position and tumor regression during stereotactic body radiotherapy for
Yanan Sun1, Yufei Lu1, Siguo Cheng2
1Department of Radiation Oncology, The Affiliated Cancer Hospital of Zhengzhou University, 127 Dongming Road, Zhengzhou, Henan 450008, China.
Purpose:
To determine interfractional changes of lung tumor centroid position and tumor regression during stereotactic body radiation therapy (SBRT).
Methods And Materials:
34 patients were treated by SBRT in 4-5 fractions to a median dose of 50 Gy. The CT scans acquired for verification were registered with simulation CT scans. The gross target volume (GTV) was contoured on all verification CT scans and compared to the initial GTV in treatment plan system.
Results:
The mean (±standard deviation, SD) three-dimension vector shift was 5.2 ± 3.1 mm. The mean (±SD) interfractional variations of tumor centroid position were -0.7 ± 4.5 mm in anterior-posterior (AP) direction, 0.2 ± 3.1 mm in superior-inferior (SI) direction, and 0.4 ± 2.4 mm in right-left (RL) direction. Large interfractional variations (≥5 mm) were observed in 5 fractions (3.3%) in RL direction, 16 fractions (10.5%) in SI direction, and 36 fractions (23.5%) in AP direction. Tumor volume did not decrease significantly during lung SBRT.
Conclusions:
Small but insignificant tumor volume regression was observed during lung SBRT. While the mean interfractional variations of tumor centroid position were minimal in three directions, variations more than 5 mm account for approximately a third of all, indicating additional margin for PTV, especially in AP direction.