Simple and effective HPLC method development and its validation for Clindipine in human drug free plasma
Selvadurai Muralidharan1, Jaya raja Kumar1, Sokkalingam Arumugam Dhanaraj1
1Faculty of Pharmacy, AIMST University, Semeling, Bedong, Malaysia.
Abstract:
Simple and effective high performance liquid chromatographic (HPLC) method was developed for estimation of Clindipine in drug free human drug free blank plasma. The internal standard used as Nifidipine (IS). The current method was used protein precipitating extraction of Clindipine from blank plasma. Separation was achieved on reversed-phase c18 column (25cm × 4.6mm, 5μ) and the detection was monitored by UV detector at 260 nm. The optimized mobile phase was used acetonitrile: 5mM potassium dihydrogen orthophosphate (pH 4.5), in the ratio of 60:40% v/v at a flow rate of 1.0 ml/min. This linearity was achieved in this method range of 10.0-125.0 ng/ml with regression coefficient range is 0.99. The present method is suitable in terms of precise, accurate and specific during the study. The simplicity of the method allows for application in laboratories that lack sophisticated analytical instruments such as LC-MS/MS or GC-MS/MS that are complicated, costly and time consuming rather than a simple HPLC-UV method. The present method was successfully applied for pharmacokinetic studies.
More Related Videos
11:29HPLC-based Assay to Monitor Extracellular Nucleotide/Nucleoside Metabolism in Human Chronic Lymphocytic Leukemia Cells
Published on: July 20, 2016
07:29HPLC Coupled with Chemical Fingerprinting for Multi-Pattern Recognition for Identifying the Authenticity of Clematidis Armandii Caulis
Published on: November 11, 2022
Related Concept Videos
Clinically Relevant Drug Product Specifications: Methods of Establishment
High-Performance Liquid Chromatography: Instrumentation