Simulation, design, and testing of a high power collimator for the RDS-112 cyclotron
Johanna L Peeples1, Matthew H Stokely1, Michael C Poorman1
1BTI Targetry LLC (formerly Bruce Technologies Inc.), 1939 Evans Rd., Cary, NC, USA.
Abstract:
A high power [F-18] fluoride target package for the RDS-112 cyclotron has been designed, tested, and commercially deployed. The upgrade includes the CF-1000 target, a 1.3kW water target with an established commercial history on RDS-111/Eclipse cyclotrons, and a redesigned collimator with improved heat rejection capabilities. Conjugate heat transfer analyses were employed to both evaluate the existing collimator capabilities and design a suitable high current replacement.