Revision of Recalled Modular Neck Rejuvenate and ABG Femoral Implants
Christopher P Walsh1, James C Hubbard2, Joseph P Nessler3
1Detroit Medical Center/Providence Hospital Orthopaedic Surgery Residency Program, Detroit, Michigan.
Abstract:
Modular neck femoral stems have been associated with adverse local tissue reactions (ALTR), leading to a voluntary recall, but these effects have not been well-characterized. A retrospective review of intraoperative findings and cobalt/chromium levels was performed in 103 hips undergoing revision for ALTR. The average preoperative serum cobalt level was 7.6 μg/L (range 1.1-23 μg/L) and chromium level was 1.8 μg/L (range 0.1-6.8 μg/L). Metallic sludge was noted in 100%, synovitis in 98%, pericapsular rind in 82%, and calcar erosion in 85%. An osteotomy was required for removal in 44%. We concluded that revision of modular neck femoral stems is associated with increased preoperative metal ion levels and stem-neck corrosion. Despite advanced stem explantation techniques, osteotomy was frequently required, leading to increased morbidity.