Quality Assurance of a 50-kV Radiotherapy Unit Using EBT3 GafChromic Film: A Feasibility Study
Femke Steenbeke1, Thierry Gevaert2, Koen Tournel1
1Department of Radiotherapy, UZ Brussel, Vrije Universiteit Brussel, Brussels, Belgium.
Purpose:
Radiochromic EBT3 film is gaining acceptance as a valuable dosimetry system for high-energy photon beams. The advantages of these films over other dosimetry systems are low spectral sensitivity and high spatial resolution. The aim of this study was to validate EBT3 film as a dosimeter for machine and treatment quality assurance (QA) of a 50-kV radiotherapy unit.
Methods And Materials:
Absolute and relative doses were acquired using EBT3 GafChromic films and compared to a parallel-plate ionization chamber (IC), the standard IC for low-energy X-rays. EBT3 was also used to evaluate beam profiles and output factors. Two films above each other, mimicking the clinical situation of a dosimeter on top of the skin, were simultaneously irradiated to evaluate EBT3 as in vivo dosimeter. All films were irradiated for 3 minutes, which corresponds with a surface dose of 3.25 ± 0.07 Gy.
Results:
A fifth-order polynomial function was found to be the best fit for the calibration curves. Good agreement between IC and EBT3 was found for absolute (0.92% for green and red color channels) and relative (1.2% and 1.0% for green and red color channels, respectively) dosimetry. Output factors for IC and EBT3 were comparable within 2.04% and 1.02% for the green and red color channels, respectively. Flatness and symmetry at the surface were within 2%. By applying film as in vivo dosimeter, an absorption of 4.70% needs to be taken into account with respect to the surface dose.
Conclusion:
EBT3 GafChromic film is a feasible and valuable QA and dosimetry tool for a 50-kV radiotherapy unit. EBT3 can be used for absolute and relative dosimetry, measurement of output factors and beam profiles. In vivo patient-specific QA can also be performed if one corrects for the dose absorption of the film.