Fracture healing using degradable magnesium fixation plates and screws
Amy Chaya1, Sayuri Yoshizawa2, Kostas Verdelis3
1Graduate Student Researcher, Center for Craniofacial Regeneration, Department of Bioengineering, Department of Oral Biology, and McGowan Institute for Regenerative Medicine, University of Pittsburgh, Pittsburgh, PA.
Purpose:
Internal bone fixation devices made with permanent metals are associated with numerous long-term complications and may require removal. We hypothesized that fixation devices made with degradable magnesium alloys could provide an ideal combination of strength and degradation, facilitating fracture fixation and healing while eliminating the need for implant removal surgery.
Materials And Methods:
Fixation plates and screws were machined from 99.9% pure magnesium and compared with titanium devices in a rabbit ulnar fracture model. Magnesium device degradation and the effect on fracture healing and bone formation were assessed after 4 weeks. Fracture healing with magnesium device fixation was compared with that of titanium devices using qualitative histologic analysis and quantitative histomorphometry.
Results:
Micro-computed tomography showed device degradation after 4 weeks in vivo. In addition, 2-dimensional micro-computed tomography slices and histologic staining showed that magnesium degradation did not inhibit fracture healing or bone formation. Histomorphology showed no difference in bone-bridging fractures fixed with magnesium and titanium devices. Interestingly, abundant new bone was formed around magnesium devices, suggesting a connection between magnesium degradation and bone formation.
Conclusion:
Our results show potential for magnesium fixation devices in a loaded fracture environment. Furthermore, these results suggest that magnesium fixation devices may enhance fracture healing by encouraging localized new bone formation.
More Related Videos
05:45Author Spotlight: Exploring Regeneration in Axolotls Through Insights in Cellular and Molecular Mechanisms, Bone Healing, and Implications for Human Therapies
Published on: April 12, 2024
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015