Related Experiment Video
Updated: Apr 18, 2026

Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
Relationship between physical factors and subjective image quality of cone-beam computed tomography images according
Jin-Woo Choi1, Sam-Sun Lee2, Soon-Chul Choi2
1Department of Oral and Maxillofacial Radiology, College of Dentistry, Dankook University, Seoul, Korea.
Objective:
This study was designed to investigate the relationship between physical factors and the subjective quality of cone beam computed tomography (CBCT) images used for different diagnostic tasks.
Study Design:
CBCT images of a real skull phantom and a SedentexCT IQ phantom were acquired under different exposure conditions (one Dinnova3 CBCT scanner, 60-110 kV and 4-10 mA). Radiologists evaluated subjective image quality of real skull phantom images for each diagnostic task. On the basis of the evaluation results, the images were classified into two groups: acceptable and unacceptable. The modulation transfer function (MTF), contrast-to-noise ratio (CNR), and image uniformity were measured using the SedentexCT IQ phantom images. The differences in physical factors were evaluated.
Results:
MTF and CNR values showed statistical differences in image quality in two groups with regard to all diagnostic tasks. In the maxilla, MTF and CNR values showed no significant differences between periapical diagnosis and implant planning in the acceptable groups. Higher MTF and CNR values were required in the periapical diagnosis compared with the implant planning of the mandible.
Conclusions:
This study proved that MTF and CNR values have a significant association with subjective image quality. The diagnostic task should be considered in evaluation of CBCT image quality.
More Related Videos
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...