Failed calibration of anesthesia machine due to look-alike oxygen sensors
Andrea P Dutoit1, Susanna R Berger, Mohanad Shukry
1From the *Department of Anesthesiology, Oklahoma University Health Sciences Center, Oklahoma City, Oklahoma; †Department of Anesthesiology, Virginia Commonwealth University, Richmond, Virginia; and ‡Department of Anesthesiology, University of Nebraska Medical Center, Omaha, Nebraska.
Abstract:
Labels and medications with similar appearances have the potential to harm patients and cause delays in hospital services. We report a problem involving the Maxtec MAX-1 and MAX-11 oxygen sensors which are commonly used on anesthesia machines. These oxygen sensors have nearly identical labels which resulted in inadvertent interchanging of the sensors. The incident required the replacement of a MAX-11 sensor with a MAX-1 sensor to ensure proper functioning of the anesthesia machine. Identification of these cases can educate health care professionals of potential sources of labeling errors and safety issues and can also bring about Food and Drug Administration policy changes.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...