Related Experiment Video
Updated: Apr 18, 2026

Use of a Filter Cartridge for Filtration of Water Samples and Extraction of Environmental DNA
Published on: November 25, 2016
Limited retention of micro-organisms using commercialized needle filters
W Elbaz1, G McCarthy2, T Mawhinney2
1Department of Bacteriology, Belfast City Hospital, Belfast Health and Social Services Trust, Belfast, UK.
Abstract:
A study was undertaken to compare a commercialized needle filter with a 0.2-μm filtered epidural set and a non-filtered standard needle. No culturable bacteria were detected following filtration through the 0.2-μm filter. Bacterial breakthrough was observed with the filtered needle (pore size 5 μm) and the non-filtered needle. Filtered systems (0.2 μm) should be employed to achieve total bacterial retention. This highlights that filtration systems with different pore sizes will have varying ability to retain bacteria. Healthcare professionals need to know what type/capability of filter is implied on labels used by manufacturers, and to assess whether the specification has the desired functionality to prevent bacterial translocation through needles.
Related Concept Videos
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Filtration
Key Techniques in Microbiology