A rapid infusion pump driven by micro electromagnetic linear actuation for pre-hospital intravenous fluid
Peng Zhao1, Yinbao Chong2, An Zhao1
1Department of Medical Engineering, Xinqiao Hospital, Third Military Medical University, Chongqing, China.
Abstract:
A rapid infusion pump with a maximum flow rate of 6 L/h was designed experimentally using a micro electromagnetic linear actuator, and its effectiveness was evaluated by comparing with that of a commercial Power Infuser under preset flow rates of 0.2, 2, and 6 L/h. The flow rate, air detection sensitivity, occlusion response time, quantitative determination of hemolysis, and power consumption of the infusion devices were extensively investigated using statistical analysis methods (p < 0.05). The experimental results revealed that the flow rate of the designed infusion pump was more stable and accurate, and the hemolysis was significantly less than that of the Power Infuser. The air detection sensitivity and the power consumption could be comparable to that of the Power Infuser except the occlusion response time. The favorable performance made the designed infusion pump a potential candidate for applications in pre-hospital fluid administration.
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...