Related Experiment Video
Updated: Apr 18, 2026

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Peripheral nerve MRI: precision and reproducibility of T2*-derived measurements at 3.0-T : a feasibility study
Alberto Tagliafico1, Bianca Bignotti, Giulio Tagliafico
1Institute of Anatomy, Department of Experimental Medicine, University of Genoa, Via L.B. Alberti, 16132, Genoa, Italy, alberto.tagliafico@unige.it.
Objective:
To prospectively evaluate the precision and reproducibility of T2*-derived measurements of the peripheral nerves.
Materials And Methods:
The study was approved by the local ethics committee and written informed consent was obtained. Bilateral upper and lower limb MRI examination was performed in 40 healthy subjects on a 3.0-T scanner. MRI protocol included T1-turbo spin-echo, T2-turbo spin-echo with fat suppression, and multiecho gradient recalled echo. Measurements of T2* times on T2* maps at different anatomical levels were performed. Three authors measured independently and in different sessions at baseline and after 4 weeks. Non-parametric tests and Bland-Altman statistics were used.
Results:
Minimum and maximum percentage variability were 10 % and 19 % for T2* (84-91 % of reproducibility). Maximum values of minimum detectable differences between limbs was 16 % (with 95 % CI: 2-37). Intra- and inter-observer agreement of the three radiologists for T2* was considered good. Evaluating the combined influence of the observer and of the repeated measurements the reproducibility was 87-98 %.
Conclusions:
T2* measurement of the peripheral nerves is precise and reproducible. The healthy contralateral side can be used as an internal control. Variations in T2* values up to 16 % have to be considered.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...