Related Experiment Video
Updated: Apr 18, 2026

Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Patient doses in paediatric interventional cardiology: impact of 3D rotational angiography
E Corredoira1, E Vañó, C Ubeda
1Medical Physics and Radiation Protection Service, Hospital Universitario La Paz, Madrid, Spain.
Abstract:
The aim of this study was to calculate the contribution of 3D rotational angiography to radiation doses received by paediatric patients in a cardiac catheterisation laboratory. The percentage increase in the median value of air kerma-area product due to cone beam CT was 33 and 16% for diagnostic and therapeutic procedures, respectively. Results are presented separately for five age groups and ten weight groups. Several methods for reducing radiation from 3D rotational angiography are suggested and patient doses are compared with previously published values.