Role of organic modifier and gradient shape in RP-HPLC separation: analysis of GCSF variants
Varsha S Joshi1, Vijesh Kumar1, Anurag S Rathore2
1Department of Chemical Engineering, Indian Institute of Technology, Hauz Khas, New Delhi 110016, India.
Abstract:
Reversed-phase high-performance liquid chromatography (RP-HPLC) of therapeutic proteins continues to play a significant role in product characterization. This study focuses on two key aspects of HPLC method development, namely the selection of organic modifier and the gradient shape. Separation of granulocyte colony-stimulating factor variants is being used as a case study to illustrate these concepts. The results demonstrate that careful selection of a binary or ternary mixture of solvents with water is an important factor to be considered for achieving the desired resolution of closely related impurities. The resolution of different types of impurities has been shown to be selective toward the choice of eluent along with the ratio in which they are mixed. In addition, this study also presents a systematic approach for selection of gradient shape based on center point solvent composition, initial solvent composition and the steepness of the gradient. The approach proposed in this study was successfully used to reduce the time of analysis from 70 min for the pertinent European Pharmacopeia method to 15 min by using a solvent system with two organic modifiers (acetonitrile and methanol) along with a sigmoidal-shaped gradient.
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
Chromatography: Introduction
The phase in which the compounds linger or on which the compounds adsorb is called the stationary phase, whereas the mobile phase is the solvent that carries the solutes to be analyzed. In traditional column chromatography, the mixture flows through the stationary phase, and the compounds partition between the stationary and mobile phases...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
High-Performance Liquid Chromatography: Instrumentation
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...