Related Experiment Video
Updated: Apr 17, 2026

Preparation and Structural Evaluation of Epithelial Cell Monolayers in a Physiologically Sized Microfluidic Culture Device
Published on: July 1, 2022
Biocompatibility of Tygon® tubing in microfluidic cell culture
Xiao Jiang1, Rex E Jeffries, Miguel A Acosta
1UNC/NCSU Joint Department of Biomedical Engineering, 4206D Engineering Building 3, 911 Oval Drive, Raleigh, NC, 27695, USA.
Abstract:
Growth of the MDA-MB-231 breast cancer cell line in microfluidic channels was inhibited when culture media was delivered to the channels via microbore Tygon® tubing. Culture media incubated within this tubing also inhibited growth of these cells in conventional 96-well plates. These detrimental effects were not due to depletion of critical nutrients due to adsorption of media components onto the tubing surface. A pH change was also ruled out as a cause. Nuclear magnetic resonance spectroscopy of the cell growth media before and after incubation in the tubing confirmed no detectable loss of media components but did detect the presence of additional unidentified signals in the aliphatic region of the spectrum. These results indicate leaching of a chemical species from microbore Tygon® tubing that can affect cell growth in microfluidic devices.