Baseline correction does not improve flow quantification in phase-contrast velocity measurement for routine clinical
Christian Meierhofer1, Christine Lyko1, Eike Philipp Schneider1
1Department of Pediatric Cardiology and Congenital Heart Disease, Deutsches Herzzentrum München, Technische Universität München (TUM), Munich, Germany.
Baseline correction using a stationary gel phantom did not improve cardiovascular magnetic resonance (CMR) flow measurements and incorrectly suggested shunts in patients without them.
Area of Science:
- Cardiovascular Imaging
- Medical Physics
- Biomedical Engineering
Background:
- Phase-contrast cardiovascular magnetic resonance (CMR) flow measurements can be affected by velocity offset errors.
- Stationary gel phantoms are proposed as a method to correct these offset errors.
Purpose of the Study:
- To evaluate the impact of baseline correction using a stationary gel phantom on flow measurement accuracy in CMR.
- To assess the effect of this correction method on shunt calculation in patients without congenital heart disease shunts.
Main Methods:
- Flow measurements were performed in 24 patients with congenital heart disease.
- Baseline correction was applied using data from a stationary gel phantom.
Main Results:
- Baseline correction led to a significant increase in incorrectly calculated shunts in patients who did not have any.
- The method did not improve the accuracy of flow measurements.
Conclusions:
- Baseline correction with a stationary gel phantom is not clinically relevant for routine CMR.
- This method does not enhance the reliability of cardiovascular flow measurements or shunt detection.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Distance Corrections
Relative Motion Analysis - Velocity
When an external force is exerted, it sets the crank into a rotational movement. This, in turn, instigates the motion of the connecting rod, leading to what is referred to as a general plane motion. This process involves two key points - point A on the connecting rod...