[Effect of oral alloy post and core on magnetic resonance imaging]
Objective:
This study aims to observe and compare the influences of the casting alloy post and cores on magnetic resonance imaging (MRI).
Methods:
Six casting alloy post and sores were fabricated for the experimental groups, and zirconia post and core served as the control. The seven types of post and cores were of the same size and shape. Each post and core was orally fastened to a volunteer and then imaged with sequences of T1-weighted spin echo imaging (SE-T1WI), T2-weighted turbo spin-echo imaging (TSE-T2WI), and diffusion weighted imaging (DWI).
Results:
Cobalt chromium alloy post and core generated severe artifacts, whereas gold palladium alloy post and core generated no obvious artifacts, and no difference was observed between the zirconia and the gold palladium alloy post and cores (P > 0.05). The pure titanium alloy post and core produced mild artifacts. The post and cores of the titanium, nickel-cadmium, and gold platinum alloys generated moderate artifacts with no significant differences (P > 0.05).
Conclusion:
Different metal post and cores have different influences on MRI. With the exception of that generated by cobalt chromium alloy, the artifact generated by the single-alloy post and core does not influence the image of the normal head and neck.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
08:36Functional Neuroimaging Using Ultrasonic Blood-brain Barrier Disruption and Manganese-enhanced MRI
Published on: July 12, 2012
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Atomic Nuclei: Magnetic Resonance
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Susceptibility and Permeability
When diamagnetic materials are placed under an external magnetic field, the moments opposite to the field are induced. Hence, the susceptibility for diamagnets has a minimal negative value of 10-5–10-6. Since...