[Properties and usage of radiophotoluminescent glass dosimeters for computed tomography dosimetry]
Ayaka Hirosawa1, Kosuke Matsubara, Hironori Kondo
1Department of Quantum Medical Technology, Graduate Course of Medical Science and Technology, Kanazawa University Graduate School of Medicine.
Abstract:
There are two types of radiophotoluminescent glass dosimeters (RPLDs). One has a tin filter in the capsule (GD-352M) and the other has no filter (GD-302M). The purpose of our study was to evaluate the properties of these RPLDs for computed tomography (CT) dosimetry: energy dependence, variation, angular dependence, and dose distribution in a single slice. Energy dependence and variation were investigated for ratio of the air kerma measured by RPLDs to that by ion chamber. Angular dependence was investigated for RPLDs and ion chamber. RPLDs were irradiated at 90°, 60°, 30°, 0°, -30°, -60°, and -90°: 0° was vertical to long axial direction of the RPLD, and plus and minus meant clockwise and anti-clockwise, respectively. Dose distribution in a single slice of an anthropomorphic phantom was acquired using 46 RPLDs. The dose responses of GD-302M and GD-352M depended on beam energy and irradiation angle of X-ray, respectively. The dose variation among dosimeters was large with GD-352M. The dose obtained from GD-352M was lower than that from GD-302M. It was expected that the dose distribution of GD-352M was formed by the primary X-ray, so RPLD without tin filter should be used for CT dosimetry.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Determination of Crystal Structures
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Photoluminescence: Applications