Biocompatibility evaluation of a thermoplastic rubber for wireless telemetric intracranial pressure sensor coating
Jun Yang1, Andrea C Charif2, Judit E Puskas2
1Department of Neurosurgery, Section of Pediatric and Congenital Neurological Surgery, CSF Physiology Laboratory, Neurological Institute, S60, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA.
Abstract:
This study investigated the biocompatibility of the experimental thermoplastic rubber Arbomatrix(™) that will be used as the protective coating on a novel intracranial pressure (ICP) sensor silicon chip. Arbomatrix(™) was benchmarked against biocompatible commercial silicone rubber shunt tubing in the brain via a rat model with 60-day implant duration. A bare silicon chip was also implanted. The results showed similar cellular distribution in the brain-implant boundary and surrounding tissues. Quantitative analysis of neuron and glia density did not show significant difference between implants. Through histological and immunohistochemical evaluation we conclude that Arbomatrix(™) is well tolerated by the brain. Due to its exceptional barrier properties Arbomatrix(™) has already been shown to be an excellent protective coating for new ICP monitoring chip.