Related Experiment Video
Updated: Apr 17, 2026

HPLC Coupled with Chemical Fingerprinting for Multi-Pattern Recognition for Identifying the Authenticity of Clematidis Armandii Caulis
Published on: November 11, 2022
Fingerprint analysis of processed Rhizoma Chuanxiong by high-performance liquid chromatography coupled with diode
Jia-Yan Fang1, Lin Zhu1, Tao Yi1
1School of Chinese Medicine, Hong Kong Baptist University, Hong Kong Special Administrative Region, Hong Kong, People's Republic of China.
Background:
Rhizoma Chuanxiong (RC) is the dried rhizome of Ligusticum chuanxiong Hort., and various types of processed Rhizoma Chuanxiong (PRC) are widely used in China. However, quality assurance and quality control of these processed medicines remain challenging. This study aims to investigate the chemical compositions of various PRC preparations by a high-performance liquid chromatography (HPLC) coupled with diode array detection (DAD) method.
Methods:
A HPLC-DAD method with validation was developed for PRC samples. Seven batches of plant samples from two processing methods, stir-frying and steaming, were analyzed by the HPLC-DAD method. Common peaks in PRC chromatograms were chosen to calculate their relative retention time (RRT) and relative peak area (RPA), and similarity analyses of the chromatographic fingerprints were conducted by Similarity Evaluation System for Chromatographic Fingerprint of Traditional Chinese Medicine software (Version 2004 A).
Results:
In the 24-h stability test, the relative standard deviation for the RRT and RPA was less than 0.07% and 2.57%, respectively. The precision was less than 0.08% for the RRT and 2.48% for the RPA. The repeatability for the RRT and RPA was less than 0.03% and 2.64%, respectively. The similarities between the seven PRC batches were range from 0.956 to 0.990. After stir-frying or steaming, the amount of ferulic acid in PRC was much higher than that in the raw material.
Conclusions:
The fingerprint analysis of PRC by different processing methods was feasible by HPLC-DAD.
More Related Videos
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Elution Process
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...