X-ray dose estimation from cathode ray tube monitors by Monte Carlo calculation
Navid Khaledi1, Azim Arbabi, Moloud Dabaghi
1*Department of Medical Radiation, Science and Research Branch, Islamic Azad University, Tehran, Iran; †Department of Medical Physics, Imam Hossein Hospital, Shahid Beheshti Medical University, Tehran, Iran; ‡Department of Medical Radiation, Science and Research Branch, Islamic Azad University, Tehran, Iran.
Cathode Ray Tube (CRT) monitors emit low-level bremsstrahlung radiation. Monte Carlo simulations calculated operator dose equivalents, finding significant reduction with leaded glass shielding.
Area of Science:
- Medical Physics
- Radiation Dosimetry
- Computational Physics
Background:
- Cathode Ray Tube (CRT) monitors can emit bremsstrahlung radiation due to electron interactions with the screen.
- Accurate dosimetry for low-dose-rate radiation from CRTs is challenging.
- Understanding operator radiation exposure is crucial for workplace safety.
Purpose of the Study:
- To calculate the dose equivalent (DE) and effective dose (ED) for operators exposed to CRT monitor radiation.
- To evaluate the protective efficacy of leaded glass shielding against CRT emissions.
- To assess ambient dose equivalent (ADE) and personal dose equivalent (PDE) for different body regions.
Main Methods:
- Utilized the Monte Carlo (MC) method with the MCNP code for radiation transport simulation.
- Simulated photon emissions from a CRT monitor and their interaction with a phantom.
- Calculated DE and ED with and without a leaded glass shield.
Main Results:
- The mean photon energy reaching the operator was calculated to be above 17 keV.
- The estimated effective dose (ED) to the phantom was 454 μSv/year (348 nSv/h).
- Leaded glass shielding reduced the phantom ED to 16 μSv/year (12 nSv/h).
Conclusions:
- CRT monitor radiation poses a quantifiable risk, necessitating protective measures.
- Conventional leaded glass significantly attenuates radiation, effectively reducing operator dose.
- The study provides valuable dosimetric data for assessing and mitigating CRT radiation exposure.