Related Experiment Video
Updated: Apr 16, 2026

Full-Circle Cauterization of Limbal Vascular Plexus for Surgically Induced Glaucoma in Rodents
Published on: February 15, 2022
Malignant glaucoma-like syndrome after EX-PRESS filtration surgery
Purpose:
To report a case of postsurgical shallow anterior chamber and elevated intraocular pressure (IOP) simulating malignant glaucoma.
Methods:
A 20-year-old woman underwent EX-PRESS® device implant for treatment of primary open-angle glaucoma.
Results:
Postoperative examination showed a shallow anterior chamber, the EX-PRESS® device embedded in the iris, an IOP of 28 mm Hg, and an annular detachment of the choroid ciliary body, suggesting hyperfiltration followed by EX-PRESS® blockage. The anterior chamber was restored using an ophthalmic viscoelastic and an additional suture was applied ensuring the scleral flap. The IOP progressively decreased in the following days and the anterior chamber remained deep and stable. New ocular ultrasonography showed complete resolution of the ciliary body detachment 15 days after surgery.
Conclusions:
After glaucoma surgery, not every patient with shallow anterior chamber and normal or high IOP necessarily has a ciliary block glaucoma. In our case, hyperfiltration led to choroidal expansion, shallow anterior chamber, obstruction of the EX-PRESS®, and secondary blockage of outflow. The differential diagnosis is important in order to avoid further invasive procedures (i.e., lensectomy or vitrectomy).
Related Concept Videos
Angle Closure Glaucoma: Treatment
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...