Ex-PRESS glaucoma filter: an MRI compatible metallic orbital foreign body imaged at 1.5 and 3T
M C Mabray1, A Uzelac1, J F Talbott1
1Department of Radiology and Biomedical Imaging, University of California, San Francisco and San Francisco General Hospital, San Francisco, CA, USA.
The Ex-PRESS glaucoma filtration device is safe for MRI scans at 1.5 T and 3 T. This small metallic implant does not cause significant artifact, ensuring diagnostic accuracy for radiologists.
Area of Science:
- Ophthalmology
- Radiology
- Medical Devices
Background:
- Glaucoma filtration devices are crucial for managing intraocular pressure.
- The Ex-PRESS glaucoma filtration device is a smaller alternative to traditional shunts.
- Understanding MRI compatibility is essential for patient care.
Purpose of the Study:
- To evaluate the MRI compatibility of the Ex-PRESS glaucoma filtration device.
- To inform the radiology community about the device's appearance and behavior during MRI.
- To assess potential risks and image interference associated with the device during MRI.
Main Methods:
- Seven patients with Ex-PRESS devices underwent MRI (1.5 T and 3 T).
- Radiology reports and electronic medical records were cross-referenced.
- MRI images were analyzed for artifact, and ophthalmology exams reviewed for complications.
Main Results:
- 18 MRI examinations were performed on 7 patients.
- No significant susceptibility artifact interfered with image interpretation.
- No complications related to MRI were observed in any patient.
Conclusions:
- The Ex-PRESS glaucoma filtration device is safe for MRI at 1.5 T and 3 T.
- The device does not produce adverse diagnostic interference due to artifact.
- Radiologists can confidently interpret MRI scans in patients with this device.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
17:16Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring
Published on: December 9, 2010
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...