Correct assignment of lipophilic dye mixtures? A case study for high-performance thin-layer chromatography-mass
Gertrud E Morlock1, Neil Brett2
1Interdisciplinary Research Center (IFZ) and Institute of Nutritional Science, Justus Liebig University Giessen, Heinrich-Buff-Ring 26-32, 35392 Giessen, Germany; University of Hohenheim, Institute of Food Chemistry, Garbenstrasse 28, 70599 Stuttgart, Germany.
Abstract:
The TLC-MS Interface, the successor of the ChromeXtract, has been available for elution head-based coupling of high-performance thin-layer chromatography with mass spectrometry (HPTLC-MS) since 2009, and is meanwhile widespread in use, mainly for compound confirmation. Until now, quantitative performance data has not been reported in detail and thus were investigated in this study. The performance data of HPTLC-electrospray ionization (ESI)-MS via the TLC-MS Interface showed good mean precisions (%RSD, n=5) for 6 dyes in a commercially available dye mixture investigated at two different concentrations (7.3% for the 1:8 dilution, and 10.1% for the 1:16 diluted) in a selected worst case scenario. The respective mean precisions of absorbance measurements were ≤1.3%. For calibrations by HPTLC-ESI-MS, the mean determination coefficient was 0.9975 for the 6 dyes (versus 0.9997 for absorbance measurement). HPTLC-MS analysis revealed the incorrect assignment of components in two commercially available dye mixtures. Using an additional software (MassWorks) that delivered a 100 times increased mass accuracy, the proposal of molecular formulae was shown to be obtainable under certain conditions with a low resolution single quadrupole mass spectrometer and in the case of helpful information such as the double bond equivalents. This enabled the identification of the incorrectly assigned unknown dyes and clearly demonstrated the benefit of using HPTLC-MS for zone confirmation.
More Related Videos
09:18Profiling the Triacylglyceride Contents in Bat Integumentary Lipids by Preparative Thin Layer Chromatography and MALDI-TOF Mass Spectrometry
Published on: September 5, 2013
09:26Cellular Lipid Extraction for Targeted Stable Isotope Dilution Liquid Chromatography-Mass Spectrometry Analysis
Published on: November 17, 2011
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Types of Detectors
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...