The Effect of Realtime Monitoring on Dose Exposure to Staff Within an Interventional Radiology Setting
Frederic Baumann1, Barry T Katzen2, Bart Carelsen3
1Miami Cardiac & Vascular Institute (MCVI), Baptist Hospital of Miami, 8900 North Kendall Drive, Miami, FL, 33176, USA. fredericbaumann@hotmail.com.
Purpose:
The purpose of this study is to evaluate a new device providing real-time monitoring on radiation exposure during fluoroscopy procedures intending to reduce radiation in an interventional radiology setting.
Materials And Methods:
In one interventional suite, a new system providing a real-time radiation dose display and five individual wireless dosimeters were installed. The five dosimeters were worn by the attending, fellow, nurse, technician, and anesthesiologist for every procedure taking place in that suite. During the first 6-week interval the dose display was off (closed phase) and activated thereafter, for a 6-week learning phase (learning phase) and a 10-week open phase (open phase). During these phases, the staff dose and the individual dose for each procedure were recorded from the wireless dosimeter and correlated with the fluoroscopy time. Further subanalysis for dose exposure included diagnostic versus interventional as well as short (<10 min) versus long (>10 min) procedures.
Results:
A total of 252 procedures were performed (n = 88 closed phase, n = 50 learning phase, n = 114 open phase). The overall mean staff dose per fluoroscopic minute was 42.79 versus 19.81 µSv/min (p < 0.05) comparing the closed and open phase. Thereby, anesthesiologists were the only individuals attaining a significant dose reduction during open phase 16.9 versus 8.86 µSv/min (p < 0.05). Furthermore, a significant reduction of total staff dose was observed for short 51 % and interventional procedures 45 % (p < 0.05, for both).
Conclusion:
A real-time qualitative display of radiation exposure may reduce team radiation dose. The process may take a few weeks during the learning phase but appears sustained, thereafter.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Biological Effects of Radiation
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...