Use of Two Intracorporeal Ventricular Assist Devices As a Total Artificial Heart
Published on: May 11, 2018
Is a Bioprosthetic Valve in the Aortic Position Desirable with a Continuous Flow LVAD?
Atsuo Doi1, Silvana F Marasco, David C McGiffin
1Department of Cardiothoracic Surgery, The Alfred Hospital, Melbourne, Australia.
Abstract:
Commissural fusion of the native aortic valve in a patient with a continuous flow left ventricular assist device (LVAD) is a known phenomenon. This may result in aortic insufficiency (AI) leading to symptomatic heart failure. In patients with AI at the time of LVAD implantation, repairing, or replacing the aortic valve is advisable. We describe a patient who had a severe dilated cardiomyopathy and moderate AI who underwent implantation of an LVAD and aortic valve replacement with a bioprosthesis that subsequently developed commissural fusion which was found at the time of heart transplantation. This case highlights the conundrum of the management of AI in patients requiring LVAD support.