Evaluation of Ignition and Burn Risk Associated with Contemporary Fiberoptic and Distal Sensor Endoscopic Technology
Kyle Spradling1, Brittany Uribe1, Zhamshid Okhunov1
1Department of Urology, University of California , Irvine, Orange, California.
Purpose:
To evaluate the ignition and burn risk associated with contemporary fiberoptic and distal sensor endoscopic technologies.
Materials And Methods:
We used new and used SCB Xenon 300 light sources to illuminate a 4.8 mm fiberoptic cable, 10 mm laparoscope, 5 mm laparoscope, rigid cystoscope, semirigid ureteroscope, flexible cystoscope, flexible fiberoptic ureteroscope, distal sensor cystoscope, and a distal sensor ureteroscope (Karl Storz, Inc., Tuttlingen, Germany). We measured peak temperatures at the distal end of each device. We then evaluated each device on a flat and folded surgical drape to establish ignition risk. Finally, we evaluated the effects of all devices on human cadaver skin covered by surgical drape.
Results:
Peak temperatures recorded for each device ranged from 26.9°C (flexible fiberoptic ureteroscope) to 194.5°C (fiberoptic cable). Drape ignition was noted when the fiberoptic cable was placed against a fold of drape. Contact with the fiberoptic cable, 10 mm laparoscope, 5 mm laparoscope, and distal sensor cystoscope resulted in cadaver skin damage. Cadaver skin damage occurred despite little or no visible change to the surgical drape. Rigid and flexible fiberoptic cystoscopes and flexible fiberoptic ureteroscopes had no effect on surgical drapes or cadaver skin.
Conclusions:
Fiberoptic light cables and some endoscopic devices have the potential to cause thermal injury and drape ignition. Thermal injury may occur without visible damage to drapes. Surgeons should remain vigilant regarding the risks associated with these devices and take necessary safety precautions to prevent patient injury.
More Related Videos
04:33Three-Dimensional Ultrasonic Needle Tip Tracking with a Fiber-Optic Ultrasound Receiver
Published on: August 21, 2018
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Procedures III: Video Capsule Endoscopy