Related Experiment Video
Updated: Apr 15, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
What a validation strategy means for the quantitation of cocaine and heroin?
Laurence Dujourdy1, Céline Charvoz2, Marion Dalmasso2
1National Forensic Institute, Central Service of Laboratories, 31 Avenue Franklin Roosevelt, 69134 Ecully Cedex, France.
Abstract:
A method of separation by gas chromatography with a flame ionisation detector was developed for quantifying cocaine and heroin in powders seized by law enforcement. The method was validated by studying parameters of calibration, trueness, precision based on trueness error (or systematic bias) and random error. Total error, which is the combination of these errors, verified its adequacy with the objectives fixed by the analyst. Accuracy profile proved to be an efficient decision tool for that purpose. Results obtained with weighted regression model were analysed and allowed to conclude that the method enables quantitation of heroin and cocaine in powders on 2-100% concentration (w/w) range with acceptance limits fixed at 10% and a risk at 5%. The possible sources of uncertainty were evaluated and measurement of their contribution was integrated. The combined standard uncertainty and expanded uncertainty were determined.